DRK310 Gas Permeability Tester- Differential Pressure Method (Single Chamber)
Short Description:
Products Gas permeability test. Suitable for O2, CO2, N2 and other non-toxic gas permeability testing of plastic film, composite film, high barrier material, sheet, metal foil, rubber and other materials Instrument Principle 1.Differential pressure method Place the pre-set sample between the high-pressure chamber and the low-pressure chamber, compress and seal, and then vacuum the high and low-pressure chambers at the same time; after vacuuming for a certain period of time and the vacuum deg...
Products
Gas permeability test.
Suitable for O2, CO2, N2 and other non-toxic gas permeability testing of plastic film, composite film, high barrier material, sheet, metal foil, rubber and other materials
Instrument Principle
1.Differential pressure method
Place the pre-set sample between the high-pressure chamber and the low-pressure chamber, compress and seal, and then vacuum the high and low-pressure chambers at the same time; after vacuuming for a certain period of time and the vacuum degree drops to the required value, close the low-pressure chamber and move to the high-pressure Fill the chamber with test gas and adjust the pressure of the high-pressure chamber to maintain a constant pressure difference on both sides of the sample; the gas permeates from the high-pressure side of the sample to the low-pressure side under the action of the pressure difference; accurately measure the pressure change in the low-pressure chamber and calculate The gas permeability performance parameters of the sample.
2. Standards compliant
YBB 00082003, GB/T 1038, ASTM D1434, ISO 2556, ISO 15105-1, JIS K7126-A.
3. Technical characteristics
Imported high-precision vacuum sensors and pressure sensors, with high test accuracy;
Semiconductor cold and hot bidirectional temperature control, parallel type, high reliability;
Precision valve piping parts, thorough sealing, high-speed vacuum, thorough desorption, reducing test errors;
Sample side leakage prevention design, reducing sample installation leakage;
Precision pressure control to maintain the pressure difference between the high and low pressure chambers in a wide range;
Patented baseline technology to remove system background leakage and reduce test errors;
Intelligent automatic: power-on self-test, to avoid failure to continue testing; one-button start fully automatic execution of the test.
Data recording: Graphical, full process and full element recording, data will not be lost after power failure.
Data security: Optional ”GMP Computerized System” software module, with user management, authority management, data audit trail and other functions.
Working environment: indoor. There is no need for a constant temperature and humidity environment (reducing the use cost), and the test data is not affected by the environmental temperature and humidity.
Parameters
Item | Parameters | Item | Parameters |
Test error | 0.01cm3/m2•day•0.1MPa | Calibrationmethod | Stand film |
Vacuum error | 0.1Pa | Vacuum range | 1333.33 Pa |
Vacuum | <10 Pa | Vacuumefficiency | lower than 27Pa in 20minutes |
Temperature range | 15℃~50℃ | Temperaturecontrol error | ±0.1℃ |
Specimenthickness | ≤3mm | Testing area | 50cm2(Circle) |
Test air | O2、N2、CO2 and non-toxic gas | Test pressure | 0.1~0.2 MPa |
Gas interface | 1/8” | Air pressure | 0.1~0.8 MPa |
Power supply | AC220V 50Hz | Power | <1500 W |
Model | DRK310 | ||
Test range | 0.1-50,000 | ||
Specimen quantity | 1 | ||
Number ofvacuum sensor | 1 | ||
Test method | Single chamber | ||
Host size | 585×640×380mm(L/W/H) | ||
Host weight | 50Kg |
The standard configuration
Test host, vacuum pump, test software, sampler, DELL computer (no BSG-11), vacuum bellows, gas cylinder pressure reducing valve and pipe fittings, sealing grease, 21.5-inch DELL display, and the host is built into the test host.
Optional accessories: container test fixture, humidity control unit.
Self-supplied parts: test gas and cylinder